Figure 7: How to hold the device with one hand (left) and two hands (right)
Hands are positioned away from the Violet Wavelength LEDs and are outside of the Illumination
Note
Zone demarcated by the raised ridge.
Figure 8: Overview of optimal distance and angling of the device relative to the wound
9.2 Handling MolecuLight i i :X
MolecuLight i:X does not require direct contact with the patient or wound during operation. The optimal distance for
imaging is 8-12 cm from the wound.
MolecuLight i:X does not require the administration of any exogenous imaging contrast agents for imaging in Fluorescence
Imaging Mode.
MolecuLight i:X can be easily cleaned between uses. Instructions for cleaning and disinfection can be found in Section 12.
MolecuLight recommends the use of standard powder-free latex or nitrile gloves when using MolecuLight i:X to reduce the
risk of bacterial contamination.
Caution
Residue from (some) medical gloves on optical components may result in distorted images.
TM
MolecuLight i:X
User Manual Revision 1.0
9.3 Imaging with MolecuLight i i :X (iX Camera App)
The MolecuLight iX Camera App is a simple and intuitive application that allows you to quickly and easily capture images
and videos on your MolecuLight i:X Imaging Device.
Warning
Caution
The iX Camera App is easily accessible by following two simple steps:
1. Push the Home Button on the MolecuLight i:X.
The MolecuLight i:X Display Screen turns on.
If the screen is locked, use 'slide to unlock' at the bottom of the display to unlock the screen.
Note
2. Press the iX Camera App icon.
The iX Camera App opens and is ready for use.
The iX Camera App allows you to:
Capture an image
Capture a 30 second video
Review previously acquired images/videos via Image Library
Delete images/videos
9.3.1 Capture an Image
The iX Camera App allows you to capture both a ST-image and FL-image by selecting the Camera icon.
9.3.1.1 Capture an ST-image
To capture an image in Standard Imaging Mode:
1. Move the Rocker Switch to the Standard Imaging Mode position.
2. Select the Camera icon.
3. When using the iX Camera App wound measurement functionality, refer to instructions provided in section 10
and Appendix E.
4. View the Range Finder LED.
Page 21 of 43
MolecuLight i:X
Only use the iX Camera App to capture wound images and/or videos.
Figure 9: Display Screen
Do not "swipe up" the camera icon circled in red above. This is not the iX Camera App that is
intended for use.
The MolecuLight i:X Imaging Device and the Display Screen must be powered on. Refer to Section
6.1 for details.
The Camera icon turns blue when selected.
TM
User Manual Revision 1.0
Page 22 of 43