6.1.1.4 Battery Status LED
The Battery Status LED indicates the MolecuLight i:X Imaging Device battery charge level.
This LED indicates the battery status for the MolecuLight i:X Imaging Device only. The Display
Screen's battery status and charging status operates independently of this functionality and is
Note
displayed on the screen.
Indicator
Solid Green Light
The MolecuLight i:X battery is fully charged or at high power (over 80% power).
The MolecuLight i:X battery power is between 20-80% power and MolecuLight i:X charging is
Solid Yellow Light
advised.
The battery power has dropped below 20% power and the MolecuLight i:X requires immediate
Solid Red Light
charging.
Table 12: Battery Status LED
When the MolecuLight i:X Imaging Device is OFF, the Battery Status LED is OFF, except when the battery is critically low and
not charging, at which point the LED blinks red every 2 seconds.
When the MolecuLight i:X Imaging Device is charging, the Battery Status LED blinks to indicate charging status of the device.
6.1.2 Rocker Switch
The Rocker Switch enables the user to toggle between Standard Imaging Mode and Fluorescence Imaging Mode.
When the Rocker Switch is in its "down" position, MolecuLight i:X is in Fluorescence Imaging Mode.
When the Rocker Switch is in its "up" position, MolecuLight i:X is in Standard Imaging Mode.
To enable Standard Imaging Mode:
1. Turn on the device (if it is not already on).
2. Toggle the Rocker Switch to its "up" position
(Figure 5).
MolecuLight i:X is in Standard Imaging Mode.
Figure 5:
Rocker Switch in Standard Imaging Mode (in 'up' position)
7 Quick Start Guide
Please see Appendix C for the MolecuLight i:X Quick Start Guide.
TM
MolecuLight i:X
User Manual Revision 1.0
Meaning
To enable Fluorescence Imaging Mode:
1. Toggle the Rocker Switch to its "down" position
(Figure 6).
MolecuLight i:X is in Fluorescence Imaging Mode.
Violet Wavelength LEDs turn on.
Figure 6:
Rocker Switch in Fluorescence Imaging Mode (in 'down' position)
Page 19 of 43
8 Environmental Conditions That Affect Use
MolecuLight i:X is intended to be used in a hospital/clinical setting.
MolecuLight i:X is not suitable for use in the presence of flammable anesthetic mixtures with air,
Caution
oxygen or nitrous oxide.
8.1 Lighting
The level of ambient light in the imaging environment dramatically affects the quality of both the Standard Imaging Mode
and Fluorescence Imaging Mode images acquired with the MolecuLight i:X Imaging Device. A green Ambient Light Status
LED indicates when the lighting conditions are optimal for imaging in Fluorescence Imaging Mode. During Standard Imaging
Mode the Ambient Light Status LED will be amber. Check the Ambient Light Status LED prior to toggling the Rocker Switch to
FL-Mode.
If you are unable to obtain optimal ambient light for Fluorescence Imaging Mode, MolecuLight recommends using
MolecuLight DarkDrape (used in conjunction with MolecuLight Adapter). For more information refer to Section 4.3.
Be aware of surroundings when using the MolecuLight i:X in dark environments during
Caution
Fluorescence Imaging Mode. Failure may result in bodily injury due to tripping.

8.2 Operating Temperature

The device is designed to operate in environment temperatures between 10-30
8.3 Storage
Store MolecuLight i:X at room temperature in a secure location between uses.
Store the MolecuLight DarkDrape at room temperature in a dry place.
9 Operating Instructions
9.1 Proper Imaging Technique
A standardized wound imaging protocol is critical to the accuracy and objectivity of digital wound evidence and image
capture. Images should be acquired while holding the device stable in the hands of the user and while the patient is not
moving (to avoid movement artifacts in the images). See Figure 7.
To capture images with MolecuLight i:X (in either Standard Imaging Mode or Fluorescence Imaging Mode):
Hold the device 8-12 cm away from the wound. The Range Finder LED is green when the distance between
MolecuLight i:X and the wound is optimal.
Ensure that the wound is centered in the camera's field of view.
Hold MolecuLight i:X such that the plane of the wound is parallel to the plane of MolecuLight i:X. See Figure 8.
The optical glass windows (see Figure 7) must be clean for optimal Standard Imaging Mode and
Note
Fluorescence Imaging Mode image capture. Refer to Section 12 for information about cleaning
MolecuLight i:X.
Caution Patient movement while the device is in close proximity to the patient may result in distorted images.
Angling the device incorrectly may result in poor image quality. The Illumination Zone on the device
Caution
should point directly at wound and be held at the recommended distance. Failure to do so may result
in poor image quality. Refer to Figure 8.
TM
MolecuLight i:X
User Manual Revision 1.0