Indications For Use; Product Specifications - apollo endosurgery ORBERA365 Manual Del Usuario

Sistema de balón intragástrico
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 186
4.

INDICATIONS FOR USE

The ORBERA365 System is to be used in conjunction with
a long-term supervised diet and behavior modification
program designed to increase the possibility of long-term
weight loss maintenance.
The ORBERA365 System is indicated for:
• Temporary use for weight loss in overweight and obese
patients (BMI 27-50) who failed to achieve and maintain
weight loss with a supervised weight-control program.
• Pre-surgical temporary use for weight loss in obese and
super obese patients (BMI 40 and above or a BMI of 35
with comorbidities) prior to obesity or other surgery, in
order to reduce surgical risk
The maximum placement period for the ORBERA365
System is 12 months, and it must be removed at that time or
earlier.
5.

PRODUCT SPECIFICATIONS

ORBERA365 System, Reference No. B-50012 (IGB
positioned in a Placement Catheter Assembly (i.e.
sheath assembly))
The IGB System contains no latex or natural rubber
materials.
The products are supplied clean, non-sterile and
packaged for single use.
The materials used to fabricate this device (see Table
1) have been tested according to ISO 10993, the
international Standard for biological evaluation of
medical devices.
Table 1: IGB Product Materials
System
Component
Silicone elastomer components coated in
IGB
Sodium Bicarbonate
Tubing:
Placement
Catheter
Assembly
Catheter Tip: Polypropylene
Sheath: Silicone elastomer and Silicone
adhesive/primer coated in Sodium
Bicarbonate
6.
CONTRAINDICATIONS
Contraindications for use of the IGB System include:
The presence of more than one IGB at the same time.
Prior surgery involving the esophagus, stomach, and
duodenum or bariatric surgery.
Any inflammatory disease of the gastrointestinal tract
including esophagitis, gastric ulceration, duodenal
ulceration, cancer or specific inflammation such as
Crohn's disease.
Potential upper gastrointestinal bleeding conditions
such as esophageal or gastric varices, congenital or
acquired intestinal telangiectasis, or other congenital
anomalies of the gastrointestinal tract such as atresias
or stenoses.
Materials
Silicone (assemblies with a PTFE
coated stainless-steel guidewire)
Polyurethane (assemblies without a
PTFE coated stainless-steel
guidewire)
A large hiatal hernia of > 5cm or a hernia ≤ 5 cm
associated
with
esophageal reflux symptoms.
A structural abnormality in the esophagus or pharynx
such as a stricture or diverticulum that could impede
passage of the delivery catheter and/or an endoscope.
Achalasia, symptoms suggestive of delayed gastric
emptying, or presence of any other severe motility
disorder that that may pose a safety risk during
placement or removal of the device.
Gastric Mass.
Severe coagulopathy.
Hepatic insufficiency or cirrhosis involving
o Acute liver failure and advanced cirrhosis with
encephalopathy muscle wasting and anasarca
o Large esophageal varices with red color signs and
gastric varices.
o Severe portal hypertensive gastropathy with or
without gastric antral vascular ectasia
Patients who are known to have or suspected to have
an allergic reaction to materials contained in the IGB.
Any other medical condition that would not permit
elective endoscopy such as poor general health or
history and/or symptoms of severe renal, hepatic,
cardiac, and/or pulmonary disease.
Serious or uncontrolled psychiatric illness or disorder
that could compromise patient understanding of or
compliance with follow up visits and removal of the
device after 12 months.
Alcoholism or drug addiction.
Patients who are unable or unwilling to take prescribed
proton pump inhibitor medication for the duration of the
device implant.
Patients unwilling to participate in an established
medically supervised diet and behavior modification
program, with routine medical follow-up.
Patients receiving aspirin, anti-inflammatory agents,
anticoagulants or other gastric irritants, not under
medical supervision.
Patients who are known to be pregnant or breast-
feeding.
7.
WARNINGS
Proper positioning of the Placement Catheter Assembly
and the IGB within the stomach (using measured
distance from the incisors via the insertion tube
markings) is necessary to allow proper filling. Lodging
of the IGB in the esophageal opening during filling may
cause serious injury. Failure to confirm proper
positioning may cause injury to the esophagus,
duodenum, or pylorus.
When filling the IGB during the placement procedure,
avoid rapid fill rates as these will generate high pressure
which can damage the IGB valve or cause premature
detachment of the IGB from the tip of the placement
catheter.
Each patient must be monitored closely during the
entire term of treatment in order to detect the
development of possible adverse events. Each patient
should be instructed regarding symptoms of deflation,
gastrointestinal obstruction, acute pancreatitis, IGB
4
severe
or
intractable
gastro-
Tabla de contenido
loading

Tabla de contenido