General Information
CAUTION!
Service connection
Access to the service menu is restricted to authorized service personnel.The con-
nected equipment must comply with the standard EN 60950 in the currently val-
id version. Do not connect a device to the service connection during surgery.
EN
CAUTION!
Peripheral devices
Additional peripheral equipment connected to interfaces of the medical monitor
has to meet the requirements of the following specifications in the respective
current valid version: IEC 60601-2-18 / EN 60601-2-18 for endoscopic devices
and IEC 60601-1 / EN 60601-1 for electrical medical devices. All configurations
have to comply with IEC 60601-1 / EN 60601-1 specifications. Whoever connects
additional equipment to signal output or signal input is considered the system
configurator and as such is responsible for complying with requirements of the
standard IEC 60601-1 / EN 60601-1.
CAUTION!
Any power cables employed by the user that are not provided by the manufac-
turer must meet the safety requirements of the national standards in the respec-
tive current valid version.
CAUTION!
Use of other accessories, other transducers and cables
The use of accessories, transducers, and cables other than those specified, with
the exception of transducers and cables sold by the MANUFACTURER of the ME
EQUIPMENT or ME SYSTEM (see Chapter 15 Glossary) as replacement parts for in-
ternal components, may result in increased EMISSIONS or decreased IMMUNITY
of the ME EQUIPMENT or ME SYSTEM.
15