Annual Inspection
EN
Fig. 10-1 Tools for pressure sensor test
Manometer
(1)
Syringe
(2)
PG012 (disposable tube set)
(3)
Silicone tube
(4)
T adapter
(5)
60
7. Stop insufflation: Press the STOP button.
The following values are displayed:
Actual pressure 0.0 (mmHg)
Gas consumption >0.0 (l)
8. Press the gas volume button.
Gas consumption 0.0 (l)
The basic function check of the device is complete.
10.3
Pressure Sensor Test (in Standard Indication and Veress Mode)
(1)
(3)
1. Set the Standard indication.
2. Select a nominal gas flow rate of 1.0 l/min.
Do not press the START/STOP button.
CAUTION!
Never use the syringe to extract gas from the device.
3. Connect a manometer and an air-filled syringe to the insufflation tube con-
nection.
4. Use the syringe to generate a pressure of at least 10 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 10 ±2 (mmHg)
5. Use the syringe to generate a pressure of at least 20 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 20 ±2 (mmHg)
6. Use the syringe to generate a pressure of at least 30 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 30 ±2 (mmHg)
10.4
Pressure Monitoring Test (in Standard Indication and Veress
Mode)
1. Set the Standard indication.
2. Select a nominal pressure of 15 mmHg and a nominal gas flow of 3 l/min.
3. Activate the Venting Valve in the user menu.
4. Use the syringe to generate a pressure of at least 27 mmHg and maximum
30 mmHg, indicated on the manometer. Start insufflation: Press the START
button. A acoustic signal is emitted (for 5 s) and the display depicts Overpres-
sure if the pressure exceeds 19 mmHg.
5. Reduce the pressure. The warning ends when the pressure falls below
19 mmHg (nominal pressure plus 4 mmHg). Stop insufflation: Press the STOP
button.
6. Select a nominal pressure of 29 mmHg.
7. Use the syringe to generate a pressure of at least 30 mmHg, which is indicat-
ed on the manometer. Start insufflation: Press the START button. An acoustic
signal is emitted with a pressure of more than 30 mmHg and the display de-
(4)