Fig. 7-5
Protocol display
The following settings can be made under indication settings:
User Menu
Setting the start pres-
sure
46
7.4
Under the "Service" menu item, the current software and firmware versions as
well as the dates of the next maintenance will be shown. Moreover, the access
for the authorized service technician is here as well.
Software: The current software version number.
Firmware: The current firmware version number.
Next maintenance: Date on which the next biannual maintenance becomes due.
Technician: password protected access for authorized service technicians. Please
consult the service manual for additional information.
7.5
Under the "Protocol" menu item, the last 10 protocols can be displayed, exported
to a memory stick, or deleted.
(2)
A list of the protocols of the last 10 surgeries is depicted. The protocols are sorted
chronologically and the protocol of the most recent surgery is listed at the top.
Select a protocol from the list to display the information about the surgery.
The following are displayed: indications used or the selected user profile; the du-
ration of surgery; the insufflation duration; smoke evac duration; minimum,
maximum and starting value for pressure and gas flow; the gas consumption;
(1)
and the gas heating details.
Use the arrow key (<-) to jump back to the protocol list.
Individual protocols can be deleted with the (-) button (see (1) Fig. 7-5 Protocol
display) or exported to a memory stick with the save button (see (2) Fig. 7-5 Pro-
tocol display). To export protocols to a memory stick, a memory stick must first
be inserted into the memory stick port Located on the device rear panel (see chap-
ter 4.4 Rear of the Device).
7.6
The following settings can be made in this menu.
Indications
Modes
Standard
Bariatric
Pediatric
Sensitive
Cardiac
Vessel Harvesting
Fieldflooding
Service
Protocol
User Menu (Overview)
Options
1-15 mmHg
1-15 mmHg
1-12 mmHg
1-12 mmHg
Pressure not adjustable
Factory Settings
15 mmHg
15 mmHg
3 mmHg
10 mmHg
Not applicable