Switching valve
Two gas bottles can be connected with the insufflator via the PG068 switching
valve. This ensures continuous insufflation, especially during replacement of the
gas bottle.
4.5.2
Connecting to Central Gas Supply
The following low pressure hoses are available for connecting to the central gas
supply:
Article
Description
PG061
Hose for CO
central gas supply DIN:US, 3 m
2
PG091
Hose for CO
central gas supply DIN:US, 5 m
2
PG092
Hose for CO
central gas supply NF:US, 5 m
2
PG093
Hose for CO
central gas supply UNI:US, 5 m
2
PG094
Hose for CO
central gas supply AGA:US, 5 m
2
PG065
Hose for CO
central gas DISS:DISS, 3.0 m
2
• Connect or loosen to the device using a 14 mm open-end wrench.
• Connect or loosen to the central gas supply receptacle on the wall.
An adapter and a low pressure hose is required for connecting to the central gas
supply with DISS connection:
• DISS Adapter for gas connection
• DISS Hose for central gas supply
1. Remove the connector (1) using 17 mm and 19 mm open-end wrenches.
2. Check if the sinter filter (3) needs a change (residues or blocked pores).
3. Exchange the sinter filter using a screwdriver.
4. Install the new connector (2) using the open-end wrenches.
For older NIST central gas supply hoses there is a NIST adapter available:
• NIST Adapter for gas connection
Set in the user menu the type of gas supply: central gas or bottle gas supply (see
chapter 7 Menu).
Initial Device Set-up
Low pressure hoses
Installation
Exchange of the gas supply adapter
EN
(3)
21