Aesculap Flow 50 Instrucciones De Manejo página 129

Idiomas disponibles
  • ES

Idiomas disponibles

The following settings can be made under indication settings:
User Menu
Indications
Gas flow rates
Standard
(depend on indica-
tion)
Keeping the (-) or (+)
button depressed for
Bariatric
1.5 seconds allows
you to select one of 3
levels for the nomi-
nal gas flow.
Pediatric
Cardiac
Venting valve
Standard
(depends on indica-
tion)
Bariatric
Pediatric
Cardiac
Max. pressure
Standard
(depends on indica-
tion)
Bariatric
Pediatric
Cardiac
Modes
Options
Veress
Gas Flow Rate 1: 1-5 l/min
Highflow
Gas Flow Rate 2: 6-30 l/min
Gas Flow Rate 3: 20-40 l/min
Veress
Gas Flow Rate 1: 1-5 l/min
Highflow
Gas Flow Rate 2: 6-40 l/min
Gas Flow Rate 3: 20-50 l/min
Sensitive
Gas Flow Rate 1: 0.1-2 l/min
Gas Flow Rate 2: 1-5 l/min
Gas Flow Rate 3: 5-20 l/min
Vessel Harvesting
Gas Flow Rate 1: 1-10 l/min
Gas Flow Rate 2: 1-10 l/min
Gas Flow Rate 3: 5-10 l/min
Fieldflooding
Not applicable
Veress
- ON
- OFF
Highflow
- ON
- OFF
Veress
- ON
- OFF
Highflow
- ON
- OFF
Sensitive
- ON
- OFF
Vessel Harvesting
- ON
- OFF
Fieldflooding
Not applicable
Veress
5-30 mmHg
Highflow
5-30 mmHg
Veress
5-30 mmHg
Highflow
5-30 mmHg
Sensitive
5-15 mmHg
Vessel Harvesting
5-20 mmHg
Fieldflooding
Not applicable
Menu
Factory Settings
Gas Flow Rate 1: 3 l/min
Gas Flow Rate 2: 20 l/min
Gas Flow Rate 3: 40 l/min
Gas Flow Rate 1: 3 l/min
Gas Flow Rate 2: 25 l/min
Gas Flow Rate 3: 50 l/min
Gas Flow Rate 1: 1 l/min
Gas Flow Rate 2: 3 l/min
Gas Flow Rate 3: 10 l/min
Gas Flow Rate 1: 2 l/min
Gas Flow Rate 2: 5 l/min
Gas Flow Rate 3: 10 l/min
OFF
ON
OFF
ON
ON
OFF
Not applicable
30 mmHg
30 mmHg
30 mmHg
30 mmHg
15 mmHg
20 mmHg
Not applicable
47
EN
loading

Este manual también es adecuado para:

Pg150