Initial Device Set-up
EN
High pressure hoses
Installation
20
of the device. This connector includes the exchangeable sinter filter and allows
the use of the different gas supply adapters described below as well.
4.5.1
Connecting a Gas Bottle
WARNING!
Keep filled CO
bottle on hand
2
Always keep a filled CO
bottle on hand ready for replacement. This avoids hav-
2
ing to interrupt surgery due to a lack of insufflation gas.
CAUTION!
Always use a high-pressure hose to connect gas bottle and device.
The gas bottle must be in a vertical position. The gas bottle pressure may not ex-
ceed 80 bar/1160.3 psi or be less than 15 bar/217.5 psi.
CAUTION!
Gas bottles with riser pipe may release dirt and oily fluids into the device. Do not
use a gas bottle with riser pipe.
The following high-pressure hoses are available:
Article
Description
PG056
High pressure hose US / bottle DIN
PG057
High pressure hose US / bottle ISO
PG058
High pressure hose US / bottle PIN index
High pressure hose with PIN connection
• Use a 14 mm open-end wrench to attach this connection on the device or to
detach it.
• The gas bottle is screwed on or off by hand.
High pressure tube with DIN connection
• Use a 14 mm open-end wrench to attach this connection on the device or to de-
tach it.
• The gas bottle is screwed on or off with a 30 mm open-end wrench.
High pressure tube with ISO connection
• Use a 14 mm open-end wrench to attach this connection on the device or to de-
tach it.
• The gas bottle is screwed on or off with a 32 mm open-end wrench.