10
Annual Inspection
Each test conducted must be documented with date and signature on the test
log.
The following measuring tools and resources were used by the manufacturer to
determine the listed measurements and tolerances:
Manometer
Range 0-100 mmHg, error class 1.6
(1)
Syringe
60 ml
(2)
PG012 (Disposable tube set)
(3)
Silicone tube
4 mm x 40 cm
(4)
(5)
T adapter
6/4/6
Veress cannula
Length 100 mm
(6)
opening diameter 1.4 mm,
inner cannula diameter 1.6 mm
An authorized service technician must check the device if the specified parame-
ters and tolerances are exceeded.
10.1
Electrical Safety Test
1. Perform a visual inspection. Make sure that
• the fuse corresponds with the specifications indicated by the manufactur-
er,
• labels and stickers on device are legible,
• the mechanical condition of the device allows for its safe use,
• the device is clean to ensure proper and safe functionality.
2. Carry out the measurements for the ground leakage current, short-circuit
current/housing leakage current, and the protective conductor resistance as
per IEC 62353 in the current version or according to the applicable national
standard.
10.2
Basic Function Test (in Standard Indication)
1. Remove insufflation tube from device.
2. Use the On/Off switch to turn the device on. The device now conducts a de-
vice check. A acoustic signal can be heard. Set the Standard operating mode.
3. A nominal pressure of 15 mmHg and a nominal flow of 3 l/min are set at the
factory.
4. The following values are displayed:
Nominal pressure 15* (mmHg)
Nominal gas flow 3* (l/min)
Actual pressure 0 (mmHg)
Gas consumption 0.0 (l)
*These values correspond with the factory setting. If values in the user menu
were changed, these changed values are displayed. Insert an PG012 (Dispos-
able tube set).
5. Start insufflation: Press the START button. The following values are displayed:
Actual pressure 0 (mmHg)
Veress Insufflation is displayed. Streaming gas can be heard at the insuffla-
tion tube.
6. Select a nominal gas flow of 20 l/min. The following value is displayed:
Nominal gas flow 20 l/min
Actual pressure 0 (mmHg)
Insufflation Active is displayed. Streaming gas can be heard at the insuffla-
tion tube.
Annual Inspection
Measured values and tolerances
EN
59