5 Technical Data

Mechanical Characteristics

Housing
Devices with a DF-1/IS-1 or DF4/IS-1 header:
Type
Connector port W x H x D in mm
VR
DF-1
65 x 55 x 12.5
DF4
65 x 52 x 12.5
DX
DF-1
65 x 55 x 12.5
DR
DF-1
65 x 55 x 12.5
DF4
65 x 56 x 12.5
HF
DF-1
65 x 58.5 x 12.5
DF4
65 x 56 x 12.5
Materials in contact with body tissue
• Housing: Titanium
• Header: Epoxy resin
• Blind plug and silicone plug: Silopren or silastic
DF4 seal: Silastic
X-ray identification
AT

Electrical Characteristics

Standards
The specifications are made according to EN 45502-2-2:2008.
Measuring conditions
If not indicated otherwise, all specifications refer to the following conditions:
• Ambient temperature: 37ºC ± 2°C
• Pacing/sensing: 500 Ω ± 1%
• Shock: 50 Ω ±1%
3
Mass g
Volume cm
36
86
34
86
36
86
36
86
35
87
37
88
36
87
20
Factory settings
• Arrhythmia zones VT1, VT2, VF: OFF
• Antibradycardia pacing: OFF
• Home Monitoring: OFF
Telemetry data
• Nominal carrier frequency: 403.6 MHz
• Maximum power of transmission: < 25 µW (-16 dBm)
International radio certification
Devices with BIOTRONIK Home Monitoring
communication.
Telemetry data for Canada and the USA:
This device must neither interfere with meteorological and earth resources technology
satellites nor with meteorological stations working in the 400,150 to 406,000 MHZ band,
and it must accept any interference received, including interference that may cause
undesired operation.
• This device will be registered with Industry Canada under the following number:
IC: 4708A-TACHNXT
The code IC in front of the certification/ registration number only indicates that the
technical requirements for Industry Canada are met.
• This device will be registered with Federal Communications Commission under the
following number:
FCC ID: QRITACHNXT
Telemetry data for Japan:
In accordance with Japanese law, this device has been assigned an identification
number under the "Ordinance concerning certification of conformity with technical
regulations etc. of specified radio equipment", Article 2-1-8.
• R: 202-SMA026
Pulse form
The pacing pulse has the following form:
The pulse amplitude reaches its maximum value at
the beginning of the pulse (Ua). With increasing
pacing duration (tb), the pulse amplitude is reduced
dependent on the pacing impedance.
®