DDDRV is the NBG code for the antibradycardia mode of the triple-chamber device:
D
Pacing in the atrium and ventricle
D
Sensing in the atrium and ventricle
D
Pulse inhibition and pulse triggering
R
Rate adaptation
V
Multisite pacing in both ventricles
VDDR is the NBG code for the antibradycardia mode of the single-chamber DX device:
V
Ventricular pacing
D
Sensing in the atrium and ventricle
D
Pulse inhibition and pulse triggering
R
Rate adaptation
VVIR is the NBG code for the antibradycardia pacing modes of the single-chamber
device:
V
Ventricular pacing
V
Sensing in the ventricle
I
Pulse inhibition in the ventricle
R
Rate adaptation
®
BIOTRONIK Home Monitoring
In addition to effective pacing therapy, BIOTRONIK provides a complete therapy
management system:
• With Home Monitoring, diagnostic and therapeutic information as well as technical
data are automatically sent to a stationary or mobile transmitter via an antenna in
the device header. The data are encrypted and sent from the transmitter to the
BIOTRONIK Service Center via the cellular phone network.
• The received data are deciphered and evaluated. Each physician can set the criteria
for evaluation to be used for each patient and can configure the time of notification
via E-mail, SMS or fax.
en • English
• A clear overview of the results of this analysis is displayed for the attending physi-
cians on the protected Internet platform Home Monitoring Service Center (HMSC).
• Data transmission from the device is performed with a daily device message.
• Device messages which indicate special events in the heart or in the device are
forwarded immediately.
• A test message can be initiated at any time using the programmer to immediately
check the Home Monitoring function.
Technical manuals
The following technical manuals provide information about usage of the device
systems:
• Technical manual for the device
• Technical manual for the HMSC
• Technical manuals for the programmer and the SafeSync Module
• Technical manual for software as online help on the user interface and as a PDF file
in the Manual Library at www.BIOTRONIK.com
• Technical manuals for the leads
• Technical manuals for cables, adapters and accessories
Order numbers for Idova with DF-1/IS-1 or DF4/IS-1 connection
Not all device types are available in every country:
Idova 7
VR-T
DF-1
383590
DF4
383591
Order numbers for Idova ProMRI with DF-1/IS-1 or DF4/IS-1 connection
Not all device types are available in every country:
Idova 7 ProMRI VR-T
DF-1
383592
DF4
383593
5
VR-T DX
DR-T
HF-T
383600
383574
383558
383575
383559
VR-T DX
DR-T
HF-T
383601
383576
383560
383577
383561