Riester RVS-100 Manual Del Usuario página 72

Ocultar thumbs Ver también para RVS-100:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 130
Alarm  range  
Average  time  
Update  Period  
PR  
Reference  method  for  the  
computation  of  PR  accuracy  
Measurement  range  
Accuracy  
Resolution  
Alarm  range  
Recovery  time  after  
PI  
defibrillation  
Measurement  range  
Masimo  SpO
 
2
SpO
 
2
A.6.3  Fast  Temp  
Measurement  range  
Sensor  type  
Measurement  range  
Resolution  
Measurement  part  
Measurement  modes  
Unit  
Accuracy  
Resolution  
Accuracy  
Average  time  
Measurement  time  
Update  Period  
Transient  response  time  
Recovery  time  after  
defibrillation  
Preheat  time  
PR  
Self-­checking  
Reference  method  for  the  
computation  of  PR  accuracy  
Alarm  range  
Measurement  range  
Alarm  indication  
Accuracy    
Resolution  
72
0%  to  100%,  high/low  limit  can  be  adjusted  continuously.  
8s,  16s  
<30s  
Electronic  pulse  simulator  
20  bpm  to  300  bpm  
20  bpm  to  250  bpm:  ±3  bpm  
251  bpm  to  300  bpm:  unspecified  
1  bpm  
0bpm~300bpm,  high/low  limit  can  be  adjusted  continuously.  
<5s  
0.05%  to  20%  
0%  to  100%  
Thermosensitive  sensor  
30.0°C~43.0°C  
1%  
Oral,  Axillary,  Rectal  
70%  to  100%:±2%(adult/pediatric,  non-­motion  conditions)  
70%  to  100%:±3%(neonate,  non-­motion  conditions)  
70%  to  100%:±3%(motion  conditions)  
Direct  mode:  Monitor  modes  
Note:  The  Masimo  SPO
sensor  for  adult/pediatric/  neonate  has  been  validated  
Adjusted  mode:  Quick  modes  and  Cold  modes  
2  
℃,  ℉  
for  motion  accuracy  in  human  blood  studies  on  healthy  adult  volunteers  in  
induced  hypoxia  studies  while  performing  rubbing  and  tapping  motions  at  2  to  
4  Hz  at  an  amplitude  of  1  to  2  cm  and  a  non-­repetitive  motion  between  1  to  5  Hz  
at  an  amplitude  of  2  to  3  cm  in  induced  hypoxia  studies  in  the  range  of  70-­
100%  SPO
 against  a  laboratory  co-­oximeter  and  ECG  monitor.  1%  has  been  
0.1℃   /  ℉  
2
added  to  the  results  to  account  for  the  effect  of  fetal  hemoglobin.  This  
variation  equals  plus  or  minus  one  standard  deviation.  Plus  or  minus  one  
Accuracy  of  Laboratory  (Constant  temperature  water  tank  ):  
standard  deviation  encompasses  68%  of  the  population.  
All  Mode  (All  Sites):  ±0.1°C  (±0.2°F)  
 
0%  to  69%,unspecified  
Adjusted  mode:  Oral  6-­10  seconds  
2-­4s,  4-­6s,  8s,  10s,  12s,  14s,  16s  
Axillary  Mode10-­14  seconds  
Rectal  Mode14-­18  seconds  
Direct  Mode  (All  Sites):  60-­120  seconds  
<30s  
<5s  
<25s(  Only  Monitor  mode)  
About  800  ms  
Every  3s  
Electronic  pulse  simulator  
30.0~43.0℃,  up-­low  range  can  be  adjustable  
25  bpm  to  240  bpm  
Three  levels  of  alarms:  sound-­light  alarms,  color  change  in  alarm  limits  area;;  and  
alarms  with  text  prompts.  
±3  bpm(non-­motion  conditions)  
±5  bpm(motion  conditions)  
1  bpm  
30
loading