uick mode is suitable for patients whose body
mperature is expected to be in the normal range
between 96.8 degrees F to 100.4 degrees F(36
egrees C to 38 degrees C).
old Preheat mode is suitable for patients whose
mperature is expected to be lower than normal
e., 91.4 degrees F, or 33 degrees C), such as
ose coming out of surgery.
onitor mode is suitable for continuous
mperature monitoring. The minimum measuring
me of this mode is recommend to be 60s.
emove the temp probe rapidly from the probe well
n the front of the monitor. This temp probe symbol
ll begin flashing as a reminder to apply a probe
ver.
disposable probe cover and position the probe
nt (see guidance below on proper positioning).
ature timer symbol
will flash while the
nt is completed.
ct mode, real-time measurement data will appear
n continuously.
measurement is completed, this probe symbol
flash as a reminder to eject the used disposable
Eject the probe cover and insert the probe back
e well.
f measurement taken without the probe cover,
ay be inaccurate.
perature Probe Positioning
rature Taking
obe tip under the tongue on one side or the other.
ent to close their mouth. Hold the probe in place
a long beep and the temperature reading is
perature Taking
ent's arm uplifted, place the probe tip into the
mpit, directly on the skin. Ask the patient to lower
d hold still. Hold the probe perpendicular to the
re is a long beep and the temperature reading is
58
Proper Temperature Probe Positioning
Oral Temperature Taking
Insert the probe tip under the tongue on one side or the other. Ask
the patient to close their mouth. Hold the probe in place until there
is a long beep and the temperature reading is displayed.
Axillary Temperature Taking
With the patient's arm uplifted, place the probe tip into the patient's
armpit, directly on the skin. Ask the patient to lower their arm and
hold still. Hold the probe perpendicular to the arm until there is a
long beep and the temperature reading is displayed.
Rectal Temperature Taking
Apply lubricant to the probe cover and insert it gently into the
patient's rectum only one-half inch to three-fourth inch (12 mm to
Rectal Temperature Taking
19 mm) for adults or one-fourth to one-half inch (6 mm to 13 mm)
Apply lubricant to the probe cover and insert it gently into the
for children. Hold the probe still until there is a long beep and the
patient's rectum only one-half inch to three-fourth inch (12
temperature reading displays.
mm to 19 mm) for adults or one-fourth to one-half inch (6 mm
to 13 mm) for children. Hold the probe still until there is a long
beep and the temperature reading displays.
Caution: If the monitor cannot take the temperature in
quick temp mode, it will automatically change modes and
Caution: If the monitor cannot take the
temperature in quick temp mode, it will
output the results. The temperature measurement site
automatically change modes and output the
and mode can only be changed when the probe is stored in
results. The temperature measurement site and mode can
its holding receptacle on the monitor. These settings can-
only be changed when the probe is stored in its holding
receptacle on the monitor. These settings cannot be changed
not be changed when the probe is out.
when the probe is out.
5.4.3 Temperature Display
5.4.3 Temperature Display
Temp value
Temp
unit of
measure
Alternate
Temp value
5.4.4 Temperature Settings
5.4.4 Temperature Settings
Select [SETTING] > [ADVANCED] > [PARAMETERS] >
1.
1.Select 【SETTINGS】 → 【ADVANCED】 → 【PARAMETERS】
[Temp] to enter the temperature setup menu.
→【Temp】to enter the temperature setup menu.
2.
Set [Unit] to [Celsius] or [Fahrenheit]. The selected measure-
ment unit will be effective during the next measurement.
2. Set【Unit】to【Celsius】or【Fahrenheit】. The selected
measurement unit will be effective during the next
measurement.
5.4.5 Safety Information
•
EU countries except Germany:
5.4.5 Safety Information
Legal regulations for monitoring instruments apply to all EU
countries except Germany.
Warning
•
Countries outside the EU:
Calibrate the device's temperature
•
measurement function every six months to one
year or according to your facility's regulations.
If the temperature exceeds the measurement
•
range, the alarm will be activated. Check whether
the temperature probe is placed on the patient's
appropriate site.
Damaged or outdated probes should be repaired or
•
replaced immediately.
13
•
•
5.5 Nurse Call
The Nurse Call function will send a signal to the nurse call system
when a patient's vital signs exceed a pre-set alarm limit. To activate
this function, the monitor must be connected to the hospital's nur-
se call system. Please use the provided the nurse-call connection
cable.
The Nurse Call function will only operate under these concurrent
conditions:
•
•
•
To set up Nurse Call:
1.
2.
3.
4.
6. Alarms
Alarms are prompts given by the monitor for medical personnel
through visual, audible and other means when either a vital sign
appears to be abnormal or a technical problem occurs.
Probe Cover
Measurement site
indicator
Upper
alarm
limit
6.1 Alarm Categories
The monitor's alarms can be classified into three categories: physi-
ologic alarms, technical alarms and prompt messages.
Lower
alarm
Physiologic alarms: Physiologic alarms are triggered by a monito-
limit
red parameter value (i.e., the DIA blood pressure value) that violates
set alarm limits. Physiologic alarm messages are displayed in the
Mode
physiologic alarm area.
Technical alarms: Technical alarms are triggered by a device mal-
function due to improper operation or system problems. The pro-
blems may result in system abnormal operation. Technical alarm
messages are displayed in the technical alarm area.
Prompt messages: As a matter of fact, prompt messages are not
alarm messages. Apart from the physiological and technical alarm
messages, the monitor will show some message to indicate the sys-
tem status.
For any countries where no legal regulations exist for monito-
ring instruments, it is recommended to examine the accuracy
of measuring instruments in 2-year intervals
If the temperature exceeds the measurement range, the
alarm will be activated. Check whether the temperature probe
is placed on the patient's appropriate site.
Damaged or outdated probes should be repaired or replaced
immediately.
The Nurse Call function is active;
An alarm condition is occurring; and
Alarms have not been paused or silenced.
Select [SETTINGS] > [ADVANCED] > [GENERAL] > [OPTI-
ONAL] and then [Enable Nurse Call]
Select [SETTINGS] > [ADVANCED] > [GENERAL] >
[ALARM] > [Nurse Call threshold] to set the alarm level at
which the nurse will be called (i.e., low, middle or high).
Select [SETTINGS] > [ADVANCED] > [GENERAL] >
[ALARM] > [Nurse Call relay type] to set the relay type to be
[Normally close] or [Normally open].
Select[SETTINGS] > [ADVANCED] > [GENERAL] >
[ALARM] > [Nurse Call trigger mode] to set the trigger
mode to be [continual] or [1s pause].
Warning: The Nurse Call function should not be used as
the primary means of patient monitoring. The care team
should evaluate alarms in combination with observations
of the patient's symptoms and overall physiological con-
dition.
Note:
•
The monitor generates all audible and visual alarms
through a speaker, LED lights and the display. When
the monitor powers on, the alarm LEDs will light once
and the speaker will beep, which indicates that the
alarm system is working properly.
•
Alarm settings are saved in real time, and then stored
in the memory of the device. After a loss of power,
the last stored settings will be shown after restarting
the monitor.
Warning: Do not set the alarm limits to extreme values
that can render the alarm system useless. Vital signs
alarm limits are pre-set by the manufacturer, but be sure
to choose clinically appropriate limits for the patient. Only
when the selected patient type is different from the last
one, the alarm limits will return to factory defaults.