SpO
 Searching  for  pulse...  
2
SpO
 Replace  Cable  
2
SpO
 Incompatible  Cable  
2
SpO
 Unrecognized  Cable  
2
SpO
 No  Sensor  
2
SpO
 Invalid  Sensor  
2
SpO
 Replace  Sensor  
2
SpO
 Calibrate  Sensor  
2
SpO
 Motion  Interference  
2
SpO
 Low  perfusion  
2
NIBP  Cuff  Type  Error  
NIBP  Air  Leak  Or  Loose  Cuff  
NIBP  Air  Pressure  Error  
NIBP  Weak  Signal  
NIBP  Range  Exceeded  
NIBP  Excessive  Motion  
NIBP  Overpressure  Detected  
NIBP  Signal  Saturated  
NIBP  Time  Out  
TEMP  No  Probe  
TEMP  too  high/  too  low  
78
The  SPO
 module  is  searching  for  pulse.  
2
The  cable  of  the  Masimo  SPO
 module  must  be  changed.  
2
The  cable  of  the  Masimo  SPO
 module  is  incompatible.  
2
The  cable  of  the  Masimo  SPO
 module  can't  be  recognized.  
2
The  sensor  of  the  Masimo  SPO2  module  can't  be  detected.  
The  sensor  of  the  Masimo  SPO2  module  is  invalid.  
The  sensor  of  the  Masimo  SPO2  module  needs  to  be  changed.  
The  Masimo  SPO2  module  is  calibrating.  
The  patient's  finger  is  moving  too  much  during  SPO2  measurement.  
The  signal  of  the  patient's  finger  is  too  low  during  SPO2  measurement.  
The  cuff  type  is  wrong.  
An  internal  valve,  air  hose,  or  the  cuff  is  leaking  air.  
The  cuff  is  not  wrapped  properly  around  the  patient's  limb.  
An  adult  cuff  is  used  in  neonate  mode.  
The  system  can't  maintain  a  stable  air  pressure.  
The  cuff  is  wrapped  too  loosely,  leading  to  a  low  patient  signal.  
The  pulse  of  the  patient  is  very  weak.  
The  NIBP  value  exceeds  the  measurement  range  (275mmHg)  
The  patient  is  moving  too  much.  
The  signal  noise  is  too  loud  during  deflation  to  detect  the  patient's  pulse  
pressure.  
The  patient's  pulse  is  random.  
There  is  too  much  cuff  pressure.  Pressure  exceeds  the  set  safe  range  (adult  
mode  is  325mmHg,neonate  mode  is  165mmHg)  
Too  much  patient  movement  has  impacted  the  NIBP  signal  amplifier.  
The  time  exceeds  120s  in  adult  mode;;  
The  time  exceeds  90s  in  neonate  mode;;  
The  fast  temp  probe  is  not  connected.  
The  temp  value  exceeds  the  measurement  range  