SpO
Searching for pulse...
2
SpO
Replace Cable
2
SpO
Incompatible Cable
2
SpO
Unrecognized Cable
2
SpO
No Sensor
2
SpO
Invalid Sensor
2
SpO
Replace Sensor
2
SpO
Calibrate Sensor
2
SpO
Motion Interference
2
SpO
Low perfusion
2
NIBP Cuff Type Error
NIBP Air Leak Or Loose Cuff
NIBP Air Pressure Error
NIBP Weak Signal
NIBP Range Exceeded
NIBP Excessive Motion
NIBP Overpressure Detected
NIBP Signal Saturated
NIBP Time Out
TEMP No Probe
TEMP too high/ too low
78
The SPO
module is searching for pulse.
2
The cable of the Masimo SPO
module must be changed.
2
The cable of the Masimo SPO
module is incompatible.
2
The cable of the Masimo SPO
module can't be recognized.
2
The sensor of the Masimo SPO2 module can't be detected.
The sensor of the Masimo SPO2 module is invalid.
The sensor of the Masimo SPO2 module needs to be changed.
The Masimo SPO2 module is calibrating.
The patient's finger is moving too much during SPO2 measurement.
The signal of the patient's finger is too low during SPO2 measurement.
The cuff type is wrong.
An internal valve, air hose, or the cuff is leaking air.
The cuff is not wrapped properly around the patient's limb.
An adult cuff is used in neonate mode.
The system can't maintain a stable air pressure.
The cuff is wrapped too loosely, leading to a low patient signal.
The pulse of the patient is very weak.
The NIBP value exceeds the measurement range (275mmHg)
The patient is moving too much.
The signal noise is too loud during deflation to detect the patient's pulse
pressure.
The patient's pulse is random.
There is too much cuff pressure. Pressure exceeds the set safe range (adult
mode is 325mmHg,neonate mode is 165mmHg)
Too much patient movement has impacted the NIBP signal amplifier.
The time exceeds 120s in adult mode;;
The time exceeds 90s in neonate mode;;
The fast temp probe is not connected.
The temp value exceeds the measurement range
37