Appendix A Product Specifications
A.1 Safety Specifications
According to the MDD 93/42/EEC, the monitor is Type Ⅱb equipment. Classified according to the IEC 60601-1 is as follows:
Classification
of protection
Parts
against
electric shock
Mainframe
I
Temp Module
NIBP Module
NA
SpO
Module
2
Note:
I: Class I, internally and externally powered equipment.
If there is doubt about protecting earth integrality or protecting the earth lead of the equipment, change the equipment to
internally powered equipment.
NA: Not applicable.
CF: Type CF applied part, defibrillation proof.
Not suitable: Equipment is not suitable for use in the presence of flammable anesthetic mixture with air or with oxygen or
nitrous oxide.
A.2 Environmental Specifications
Operating temperature
Operating humidity
Operating atmospheric pressure
Transportation and storage
temperature
Transportation and storage
humidity
Transportation and storage
atmospheric pressure
A.3 Physical Specifications
Parts
Weight ( kg )
Mainframe
A.4 Power Specifications
Input voltage
Frequency
66
Degree of
protection
against electric
shock
No mark
Type CF applied
part defibrillation
proof
+5℃ to +40℃
15% to 85%(non condensing)
700hPa to 1060hPa
-20℃ to +55℃
10% to 93%(non condensing)
500hPa to 1060hPa
Size(W×H×D)(mm)
<4kg
314mm × 132mm × 239mm
100V-240V AC
50Hz/60Hz
Degree of
Degree of
protection
protection
against
against
ingress of
hazards of
liquid
explosion
IPX1
Not suitable
Remarks
Including screen, stationary parameter
module, a lithium battery, without
accessories.
24
Mode of
operation
Continuous