14. At 25 Ohms, observe the maximum power reading at 50 Watts. <RECORD RESULTS>
15. Stop the test on the ESuA.
16. Set the load of the ESuA to 50 Ohms
17. Start the test on the ESuA.
18. Press and hold the activation button.
19. At 50 Ohms, observe the maximum power reading at 25 Watts.<RECORD RESULTS>
20. Press and hold the activation button.
21. At 50 Ohms, observe the maximum power reading at 35 Watts. <RECORD RESULTS>
22. Press and hold the activation button.
23. At 50 Ohms, observe the maximum power reading at 50 Watts. <RECORD RESULTS>
24. Stop the test on the ESuA.
25. Set the load of the ESuA to 225 Ohms
26. Start the test on the ESuA.
27. Press and hold the activation button.
28. At 225 Ohms, observe the maximum power reading at 25 Watts. <RECORD RESULTS>
29. Press and hold the activation button.
30. At 225 Ohms, observe the maximum power reading at 35 Watts.<RECORD RESULTS>
31. Press and hold the activation button.
32. At 225 Ohms, observe the maximum power reading at 50 Watts. <RECORD RESULTS>
33. Stop the test on the ESuA.
34. Power everything down and disconnect the UUT from the test equipment.