System Error
Instrument Error
Bolder Surgical
• A single-pulsed alert tone
• RF energy delivery stops
Power button illuminates
and remains red
• A single-pulsed alert tone
• Instrument receptacle display
illuminates amber and flashes
continuously
• Generator will not allow RF
energy delivery
Generator not functioning properly,
failed an internal self-check.
Generator not functioning properly,
failed an internal self-check.
Possible use conditions include:
Inadvertent depression of the
instrument activation button during
instrument connection.
The instrument switch is
malfunctioning.
Page | 15
1. Press the power button to turn
off the generator
2. Wait at least 30 seconds
3. Press the power button to turn
on the generator
4. Confirm the power button
illuminated green before use.
If the system error reoccurs:
• Do not use the generator.
• Use a different CoolSeal
Generator or alternate means
to complete the procedure.
Return the malfunctioning
generator as instructed in
the Product Service section
of this guide.
1. Disconnect instrument from
the generator.
2. Ensure instrument activation
button is not being pressed.
3. Reconnect instrument to the
generator.
4. Confirm the instrument
receptacle display
illuminates green
If the Instrument Error reoccurs:
• Do not use the instrument
• Use a different CoolSeal
Instrument.
Remove anything depressing the
instrument activation button and
reconnect instrument.
Replace the instrument.