EARTH LEAKAGE TESTING
1.
Turn on the ESA.
2.
Plug the power cable from the UUT into the ESA.
3.
Insert the Functional Test Cable into the instrument receptacle.
4.
Set the Functional Test Cable to Normal, stack the RED and BLACK banana connectors
5.
Insert the stacked connectors into the RA/R (resistance) port of the ESA
6.
Using the Micro Amp setting (µA) and EARTH, perform the following series of tests.
7.
Normal/Closed <RECORD RESULTS>
Normal/Open <RECORD RESULTS>
8.
Reverse/Open <RECORD RESULTS>
9.
10. Reverse/Closed <RECORD RESULTS>
11. Remove the Functional Test Cable and repeat the above series of tests.
12. Normal/Closed <RECORD RESULTS>
13. Normal/Open <RECORD RESULTS>
14. Reverse/Open <RECORD RESULTS>
15. Reverse/Closed <RECORD RESULTS>
16. Disconnect the UUT from the ESA, power off the ESA.
IMPEDANCE TESTER
1.
Turn on the ESuA
2.
Power on the UUT
3.
Set the Functional Test Cable switch to Normal
4.
Plug the RED and BLACK Functional Test Cable plugs into the corresponding ESuA VAR LOAD sockets.
5.
Insert the Functional Test Cable into the UUT
6.
Confirm that the Power Button is GREEN, and the instrument receptacle display is WHITE before continuing.
Important
The CoolSeal
™
second activation is at a power output of 35 Watts and the third activation is at a power output of 50 Watts. The
fourth activation begins this sequence again. To restart the sequence from a known starting point, simply unplug
the Functional Test Cable and reinsert it.
Important
When using the Functional Test Cable the CoolSeal
changes significantly. This is unlikely to occur during the Functional Test, but if it were to happen, simply power the
unit off and on again to reset the unit.
7.
Set the load of the ESuA to 25 Ohms.
8.
Start the test on the ESuA.
9.
Press and hold the activation button.
10. At 25 Ohms, observe the maximum power reading at 25 Watts. <RECORD RESULTS>
11. Press and hold the activation button.
12. At 25 Ohms, observe the maximum power reading at 35 Watts. <RECORD RESULTS>
Bolder Surgical
Generator tests a range of power outputs. The first activation is at a power output of 25 Watts, the
Generator will put the unit into a failure state if the resistive load
™
Page | 34
en - CoolSeal™ Generator