Caution
Set the system volume control at a level that ensures the activation tones can be heard.
Do not turn the activation tone down to an inaudible level. The activation tone alerts the surgical team when the
CoolSeal
Generator is delivering RF energy.
™
Examine all instruments and connections to the system before using. Ensure that the instruments function as
intended. Improper connection may result in arcs, sparks, instrument malfunction, or unintended surgical effects.
A non-functioning CoolSeal
for use.
The use of appliance couplers, i.e., power cords, other than those specified may result in increased emissions or
decreased immunity of the equipment.
Provide as much distance as possible between the CoolSeal
monitors. Do not cross or bundle electronic devices cords. This generator may cause interference with other
electronic equipment. If interference occurs, relocate the equipment.
Studies have shown that smoke generated during electrosurgical procedures can be potentially harmful to
patients and the surgical team. These studies recommend adequately ventilating the smoke by using a surgical
smoke evacuator or other means.
1.
U.S. Department of Health and Human Services. National Institute for Occupational Safety and Health (NIOSH). Control of Smoke from Laser/Electric
Surgical Procedures. HAZARD CONTROLS, Publication No. 96-128, September 1996.
Setting up the CoolSeal
1.
Place the CoolSea
non-conductive wheels are recommended. Refer to the procedures for your local institution or your local codes.
2.
If required by institution or local codes, connect the equipotential connector to a ground source.
3.
Connect the system power cord to the electrical inlet on the rear panel.
4.
Connect the system power cord to a grounded AC wall receptacle.
Notice
Do not connect the CoolSeal
Do not position the device so that it is difficult to disconnect the power cord.
Important
If transported or stored at temperatures outside the operating temperature range, allow one hour for the CoolSeal
Generator to reach operating temperature before use.
Turn on the system by pressing the power stand-by button (
power-up self-test:
•
Unit performs start-up self-test.
•
The power button illuminates green.
•
Audible alert sounds three (3) ½ second tones after successful completion of self-check.
Notice
Verify that the system has successfully completed the start-up self-test before connecting instruments.
If unit fails start-up self-test:
•
Power button does not illuminate or illuminates red
•
Audible alert tone sounds one (1) ½ second tone after unsuccessful self-check
Refer to the Alert Conditions or Troubleshooting sections for additional information
Bolder Surgical
Generator may cause interruption of surgery. A backup system should be available
™
1
Generator
™
Generator on a flat, stable surface such as a table, platform, boom system, or cart. Carts with
l™
Generator to a power strip or extension cord.
™
Generator and other electronic equipment such as
™
) on the front of the unit. Observe the following during the
Page | 12
™
en - CoolSeal™ Generator