Chapter 3. Setup and Use of the CoolSeal
This chapter covers the following topics:
•
Unpacking the CoolSeal
•
Inspecting the CoolSeal
•
Setup Safety
•
Setting up the CoolSeal
•
Adjusting the volume of activation tones
•
Connecting a CoolSeal
•
Activating the CoolSeal
•
Descriptions of Alert Conditions
Caution
Read all warnings, cautions, and instructions provided with this generator before use. Specific instructions for
CoolSeal
instruments are not included in this manual. Refer to the CoolSeal
™
warnings, cautions, and instructions for its use.
Unpacking the CoolSeal
1.
Remove the CoolSeal
2.
Inspect the device for visible damage.
Inspecting the CoolSeal
Before each use of the CoolSeal
•
Inspect for damage to the Generator and all its connections.
•
Inspect all cords and connectors for signs of wear, damage, and abrasion.
•
Verify that no errors occur when you turn on the unit.
Setup Safety
Warning
Electric Shock Hazard: Connect the system power cord to a properly grounded power receptacle. Do not use power
plug adapters.
Fire Hazard: Do not use extension cords or multiple socket outlets.
Patient Safety: Use the CoolSeal
manual, otherwise inaccurate power outputs may result.
This equipment is for use only by trained, licensed physicians. Use of this equipment without such training can result
in serious patient injury.
The CoolSeal
™
Do not wrap the instrument cords around metal objects. This may induce stray currents that could lead to shocks,
fires, or injury to the patient or surgical team.
Electric Shock Hazard: Do not connect wet instrument connectors to the CoolSeal
instruments and adapters are correctly connected.
Inspect equipment cable for breaks, cracks, nicks and other damage before every use. If damaged, do not use.
Damaged equipment cables may result in injury or electrical shock to the patient or surgical team.
Bolder Surgical
Generator
™
Generator
™
Generator
™
Instrument
™
Instrument
™
Generator
™
Generator from the shipping box and packaging material.
™
Generator
™
Generator, verify that the unit and all accessories are in good working order:
™
Generator only if the power-up self-test has been completed as described in this
™
Generator is only compatible with CoolSeal
Generator
™
Instruments. No other bipolar instruments can be used.
™
Page | 11
instrument's instructions for use for
™
Generator. Ensure that all
™
en - CoolSeal™ Generator