Adjusting the volume

The volume of alert and activations tones may be adjusted by turning the volume knob on the back panel of the
CoolSeal
Generator.
Notice
The minimum volume for alert tones will always be louder than the activation tone.
Connecting a CoolSeal
Warning
Electric Shock Hazard
Do not connect wet instruments to the CoolSeal
Ensure that all instruments are correctly connected and that no metal is exposed at any connection point.
Properly seat the instrument in the receptacle. Improper connection may result in inadvertent instrument activation
or other potentially hazardous conditions. Follow the instructions provided with CoolSeal
connection and use.
The instrument receptacle on this system is designed to accept only one instrument at a time. Do not attempt to
connect more than one instrument at a time into a given receptacle. Doing so will cause simultaneous activation
of the instruments.
Caution
Inspect instruments and cords for breaks, cracks, nicks and other damage before every use. If damaged, do not use.
Damaged instruments or cords may result in injury or electrical shock to the patient or surgical team.
Connecting a CoolSeal
1.
Connect a CoolSeal
2.
Verify proper placement and connection by observing the green lit instrument receptacle display.
If the CoolSeal
Generator does not recognize the attached instrument, the instrument receptacle display will Illuminate red or not
illuminate and a single alert tone will sound. Refer to the Alert Conditions or Troubleshooting sections.
Activating a CoolSeal
1.
Activate the CoolSeal
2.
Observe the following during instrument activation:
The activation display illuminates blue and varies in brightness for the duration of energy delivery.
An activation tone sounds continuously during energy delivery.
3.
When the seal cycle is complete, observe the following:
A three-toned sequence will sound when the seal is completed. The activation button should then be released.
The activation display on the CoolSeal
The activation display will then extinguish.
In the case of an alert condition, refer to the Alert Conditions or Troubleshooting sections.
Bolder Surgical
Instrument
Instrument to the CoolSeal
Instrument to the instrument receptacle on the front panel of the CoolSeal
Instrument
Instrument per the Instrument's instructions for use.
Generator front panel will momentarily brighten blue.
Generator.
Page | 13
Instruments for proper
Generator
Generator.
en - CoolSeal™ Generator
Tabla de contenido
loading

Este manual también es adecuado para:

Csl-200-50

Tabla de contenido