Invalid Instrument
Bolder Surgical
• A single-pulsed alert tone
• Instrument receptacle display
illuminates and remains red
• Generator will not allow
RF energy delivery
An unusable instrument has been
connected.
Possible use conditions include:
Instrument connected is not
CoolSeal
technology compatible.
Instrument connected has been
used previously.
Instrument is not usable with
software version.
Page | 16
1. Disconnect instrument from
the generator.
2. Ensure instrument is
CoolSeal
compatible
3. Reconnect instrument to the
generator.
4. Confirm the instrument
receptacle display
illuminates green
If the Instrument Error reoccurs:
• Do not use the instrument
• Use a different CoolSeal
Instrument.
Confirm CoolSeal
technology
compatibility of the instrument
from the instrument instructions
for use.
Discard instrument.
Confirm the software version
required by the instrument is not
greater than the software version
labelled on the bottom of the
CoolSeal
Generator.
For a software upgrade refer to
the Software Upgrade section.