5.1 Safety checks
The safety check must be carried out once a year.
Note:
The result of the safety check should be recorded in a test report, see Chapter 7. Appendix.
Work to be carried out
Visual inspection
– Housing and accessories (socket modules,
frame)
– Inscriptions, manufacturer's identification
plate data
– CE mark, Storz inspection label
– Instruction Manual
– Power fuses
– Fuse labels
– Internal parts of the device
– Feedthrough covers
– Software versions
Test for proper operation
– Power switch
– Device startup
Touch screen pressed
Footswitch pressed
Handpiece button pressed
– Footswitch detection
– Activation signal (bipolar Cut, bipolar Coag)
– Switching between procedures A / B (triple-
pedal footswitch only)
– Handpiece detection activation signal (Cut,
Coag)
– NE monitoring critical resistance (single-
faced / double-faced)
– NE monitoring symmetry (double-faced)
– Autostart / autostop (bipolar)
– Instrument detection MF socket
– Power measurement, monopolar cut
– Power measurement, bipolar cut
– Power measurement, monopolar coag.
– Power measurement, bipolar coagulation
– KARL STORZ-SCB
1) For test for proper operation, please see Chapter 4 Testing and Adjustments.
5-2
1)
®
bus termination
Remarks
No external damage
Correct, legible, clean, wipeable, securely attached
Fixed to the housing
Present
Correct values, undamaged, secure,
2 x T 8.0 AH 250V (100 V ... 120 V)
2 x T 4.0 AH 250 V (220 V ... 240 V)
Affixed above the fuse holder, correct fuse ratings
Free from visible faults
Present
Generator control, ESI, user interface,
KARL STORZ-SCB
®
Lights up green when on
Procedure list, alive signal (clock)
Error message
Error message
Error message
Power Cut
300W ±20%
Top Cut
300W ±20%
C-Cut
200W ±20%
Bipolar Cut
100 W ±20%
Saline C-Cut (MF) 370 W +9% / -20%
Standard Coag
200 W ±20%
Forced Coag
120 W ±20%
Spray Coag
120 W ±20%
Bipolar Soft Coag 120 W ±20%
Saline Coag (MF)