Calibrating - Storz AUTOCON II 400 Manual De Servicio

– Confirm the new service password by entering it again in the input field on the right.
– Press "Enter"

4.5.6 Calibrating

Caution: The calibration process must be carried out with extreme care and precision.
– Open the "Service Programs" menu as described in Chapter 4.5.1 Calling the service programs.
– Press the "Calibration" button.
– Four register marks will appear in succession on the screen. You should touch them in the center
using a pencil / (ballpoint) pen. Once the last register mark has been pressed, the menu will jump back
to the "Service Programs" screen.
The calibration has been accepted automatically.
4-18
to confirm the new service password.
If one of the register marks is actuated imprecisely, then the calibration data
from the touch screen will be lost.
There is no facility for recalibrating at this point, and in this event you should
contact the service technician.
Figure 4.17
Figure 4.18
AUTOCON
®
II 400
MODEL 205352 20-1
12.06 / V1.0