Storz AUTOCON II 400 Manual De Servicio página 169

Error code
Description of error
I-81
Short circuit in instrument interface
I-82
Short circuit in activation recognition
I-83
Cut-off in activation recognition
I-84
Instrument interface: R-Window
I-85
Short circuit in activation recognition
I-86
Illegal activation type
I-87
Framing error
I-88
Parity error
I-89
Unknown ECB identifier
I-8A
CRC error in instrument recognition
I-8B
Instrument interface: R_S3
Non permissible socket module
I-8C
recognized
I-8D
Instrument interface: R_S4
I-8E
Instrument interface: R_S8
I-90
EEPROM write error
N-01
Interrupt error
N-02
CRC error following a reset
N-03
+5 V too low
N-04
+5 V too high
N-05
Undervoltage of ±12 V
N-06
Overvoltage of ±12 V
N-07
Current for NE monitor absent
N-08
SPI receive error
N-09
Stack error
N-0D
Error in state variable
N-0E
CRC error during operation
N-10
Defective current sensor 1
N-11
Defective current sensor 2
Caution: Always unplug the device from the power supply before carrying out any
12.06 / V1.0
repairs on it!
Solution
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Check connected instrument and
replace if necessary
Arrange for Black Box to be tested and
replaced if necessary
Check connected instrument and
replace if necessary
Check connected instrument and
replace if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
Arrange for Black Box to be tested and
replaced if necessary
AUTOCON
®
II 400
MODEL 205352 20-1
Priority
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
Low
3-11

Hide quick links:

Capítulos

Solución de problemas

loading