Measuring Power - Storz AUTOCON II 400 Manual De Servicio

4.3.15 Measuring power

Note:
In order to measure the power, both the device and the measuring equipment must be
operated in Standby mode for at least 30 mins. beforehand.
– Connect the device to the correct supply voltage.
– Connect the dual-pedal footswitch 200138 30.
– Start the device.
– Set the NE to the value "any".
– Select any procedure.
– Set the modes, the power settings and effects in accordance with the test report.
– Use the appropriate test cables for the HF socket modules according to the device version:
Example 205352 20-115: Measuring cable with 4 mm lab socket for
bipolar round
MF socket
monopolar socket
NE jack
– The basic test configuration is represented in the Figure 4.9 below.
STORZ
autocon II 400
12.06 / V1.0
205352 20
F
Figure 4.9
AUTOCON
MODEL 205352 20-1
Metron QAES
F1
F2
F3 F4
F5
Foot
switch
Scope
connections monopolar
connections bipolar
connections multifunctional (MF)
®
II 400
Fixed
Var.
load
load
4-11

Hide quick links:

Capítulos

Solución de problemas

loading