The separate modes are indicated using numbers. These should be assigned as follows:
Number
Mode (Abbreviation)
0
HFM_BI_CUT_OFF
1
HFM_BI_CUT_BIPOLAR
10
HFM_MO_CUT_OFF
11
HFM_MO_CUT_POWER
12
HFM_MO_CUT_TOP
13
HFM_MO_CUT_LAP_C
14
HFM_MO_CUT_C
15
HFM_MO_CUT_GASTRO
16
HFM_MO_CUT_PAPILLO
20
HFM_MF_CUT_OFF
21
HFM_MF_CUT_SALINE_C
22
HFM_MF_CUT_SALINE_TIME_C
100
HFM_BI_COAG_OFF
101
HFM_BI_COAG_SOFT
102
HFM_BI_COAG_SOFT_AUTO_STOP
110
HFM_MO_COAG_OFF
111
HFM_MO_COAG_STANDARD
112
HFM_MO_COAG_FORCED
113
HFM_MO_COAG_SPRAY
120
HFM_MF_COAG_OFF
121
HFM_MF_COAG_VASCULAR_SAFE
122
HFM_MF_COAG_SALINE_COAG
123
HFM_MF_COAG_SALINE_TIME
4.5.3 Resetting the safety checks
– Open the "Service Programs" menu as described in Chapter 4.5.1 Calling the service programs.
– Press the "Reset Date Safety Test" button.
– Press "Enter"
– Switch the device off.
– Turn the device back on.
The device is switched on without any alarm messages.
4.5.4 Output relays on / off
– Open the "Service Programs" menu as described in Chapter 4.5.1 Calling the service programs.
– Press the "Output relays on / off" button.
The output relays are closed or opened.
Note:
The closed relays are acknowledged by a green dot on the button.
4-16
to confirm the input.
Mode
Cut off
Bipolar Cut
Cut off
Power Cut
Top Cut
Lap C-Cut
C-Cut
Gastro Cut
Papillo Cut
Cut off
Saline C-Cut
Saline Time C-Cut
Coag off
Bipolar Soft Coag
Bipolar Soft Auto Stop
Coag off
Standard Coag
Forced Coag
Spray Coag
Coag off
Bivascular Safe
Saline Coag
Saline-Time-Coag
AUTOCON
®
II 400
MODEL 205352 20-1
Socket
Bipolar
Bipolar
Monopolar
Monopolar
Monopolar
Monopolar
Monopolar
Monopolar
Monopolar
Multi-function
Multi-function
Multi-function
Bipolar
Bipolar
Bipolar
Monopolar
Monopolar
Monopolar
Monopolar
Multi-function
Multi-function
Multi-function
Multi-function
12.06 / V1.0