80
Technische Beschreibung
Fehlersuche
Error code
Symptom
A-0B:
NE contact: Resistance too low
A-0C:
NE contact: Resistance too high
A-10:
End activation (AutoStop reached)
A-11:
No activation signal
A-12:
Activation during power up not permitted
A-17:
Simultaneous activation not permitted
A-19:
Line voltage fault present
A-1C:
Max. activation time exceeded
A-1E:
Touch screen pressed during power up
A-1F:
NE symmetry critical (cancel activation)
A-22:
AutoStop reached
A-85:
Dual-pedal footswitch disconnected
A-89:
Single-pedal footswitch disconnected
A-94:
Triple-pedal footswitch disconnected
A-95:
Instrument detected at MF socket
A-9F:
MF instrument disconnected
A-A3:
Footswitch not assigned to a socket
A-A4:
2 footswitches of same type connected
A-A9:
First select procedure before activation
A-AA
No mode assigned
A-AB
No/Unsuitable MF instrument detected
A-AC
Autostart on tissue contact not possible
A-B0
NE symmetry not optimal
A-B7
Autostart only permitted up to 50W output
A-BB
SC due
A-C0
Activation type for MF-instrument missing
A-FE
Line voltage fault present
C-84
Short-circuit between AE and NE
all others
Technical description
Troubleshooting
AUTOCON
®
II 400 Error Messages
Troubleshooting
Check neutral electrode and reposition if necessary
Check neutral electrode and reposition if necessary
End activation of HF energy
Have unit serviced.
Check footswitch, handle and cable
Activate either via footswitch or handle
Check power cord, power outlet and power supply
Cancel activation
Touch screen must not be pressed during power-up
Check neutral electrode and reposition if necessary
Break contact with tissue
Reconnect dual-pedal footswitch and check cable
Reconnect single-pedal footswitch and check cable
Reconnect triple-pedal footswitch and check cable
New MF instrument at socket
MF instrument removed from socket
Assign footswitch to a socket (p. 56)
Either connect one footswitch or two of a different type
Select the required procedure (p. 52)
Assign a mode to the procedure (p. 53)
Attempt to activate the socket where there was no or a defective MF
instrument detected.
Release tissue contact, separate instrument tips, if necessary, assign
Autostart
Check neutral electrode and reposition if necessary
The max. output with Autostart should only be 50 W
Arrange date for safety check
Specify activation type
Check power cord, power outlet and power supply
No impedance modification of tissue during coagulation
Inform Service
Descripción técnica
Localización de errores
Priority
Alarm condition
Medium
physiological
Medium
physiological
Low
other
High
technical
Medium
other
Medium
other
Low
technical
Low
other
Low
technical
High
technical
Low
other
Low
technical
Low
technical
Low
technical
Low
technical
Low
technical
Low
other
Low
other
Low
other
Low
other
Medium
technical
Medium
other
Low
technical
Low
technical
Medium
technical
Low
other
Low
technical
Low
technical
Low