before use
the precautionary measures listed below taken. Please read
about additional procedures and precautionary measures in
appropriate medical publications.
•
Reocor D may only be used by persons with knowledge of
cardiology who were trained in the handling of the device.
Potential users are technical and medical hospital staff
and physicians.
•
Reocor D interacts with the human heart. There is also
an interaction with the patient's skin and blood vessels.
•
Reocor D and the cables and accessories approved along
with the device may only be used in accordance with this
technical manual.
•
Reocor D must not be connected to other electromedical
devices.
•
Reocor D must not be used in areas in which there is a
danger of explosion.
•
Only the manufacturer or a party expressly authorized by
BIOTRONIK may perform corrective maintenance,
enhancements or modifications to the device.
•
To ensure safety compliance, use only original replace-
ment parts and accessories authorized by BIOTRONIK.
Using any other parts voids the manufacturer's liability
for any consequences, guarantee and warranty.
•
In case of pacemaker dependency of the patient, an
emergency pacemaker should be kept on hand.
•
Keep an external defibrillator, oxygen, intubation equip-
ment and emergency drugs on hand.
•
Before use, Reocor D should be visually inspected for
damages and dirt.
•
Never use a device that is damaged or shows abnormal
behavior. Replace any cable that shows even slight
damage.
•
Before using Reocor D, the patient cable or leads, the
user should touch the patient to equalize electrical
potentials.
•
It is strongly recommended that users examine all set
parameters before the leads are connected to Reocor D.
•
Even though Reocor D is protected from dripping water,
the device and all plugs should be kept clean and dry.
•
Reocor D cannot be sterilized.