30
Symbols
IP31
OFF
Adjustable parameters
Pacing modes
Basic rate
Pulse amplitude (A, V)
Sensitivity (A)
Sensitivity (V)
AV delay
Burst rate (A)
Fixed parameters
Pulse width
Auto short after pace
Interference interval
In channel blanking
Cross channel blanking
Total atrial refractory
period (TARP)
TARP minimum
for (30 ... 120) ppm
for (121 ... 250) ppm
Technical Data
Follow the instructions for use in the technical manual.
Indication of the placement of the battery in the compartment
Disposal according to the WEEE directive
Classification: CF (cardiac floating) applied part, defibrillation
protected
Water-repellent, protection degree IP31
Off (on the Mode dial)
DDD, D00, VDD, VVI, V00, VVT
(30 ... 250 ppm) ± 1 ppm
0.1 ... 17 V ± max.
(50 mV, 10%)
0.2 ... 10 mV ± 15%
1 ... 20 mV ± 15%
(15 ... 400 ms) ± 4 ms
(60 ... 1000 ppm) ± 20 ppm
1 ms ± 5%
< 20 ms ± 10%
80 ms ± 5 ms
110 ms ± 3 ms
19 ms ± 3 ms
AVD + 175 ms ± 5 ms
400 ms ± 5 ms
240 ms ± 5 ms
At a rate of > 180 ppm a warning
tone is emitted
At a pulse amplitude of < 1 V a
warning tone is emitted
With respect to 15 ms sin
With respect to 40 ms sin