12
Designation
1a
Atrial channel connection
1b
Ventricular channel connec-
tion
2
Battery compartment
3
Low battery LED
4
Ventricle Pace LED
5
Ventricle Sense LED
6
Ventricle Ampl. control dial
7
Ventricle Sens. control dial
8
Atrium Pace LED
9
Atrium Sense LED
10
Atrium Ampl. control dial
11
Atrium Sens. control dial
12
Start burst
13
Velcro harness and hanger
(on back)
14
Burst rate atrium control dial Setting the atrial burst rate
15
AV delay control dial
16
Select burst
17
Mode dial
18
Rate control dial
Table 1: Description of elements in Figure 1
Function
For cables and leads with 2-mm plugs or for Redel adapters
(red = plus; blue = minus)
For 9-V block battery
Provides a warning when the battery voltage is too low
Yellow display for ventricular stimulated event
Green display for sensed R wave
Setting the ventricular pulse amplitude
Setting the ventricular sensitivity
(cannot be used in D00 and V00 operating modes)
Yellow display for atrial stimulated event
Green display for sensed P wave
Setting the atrial pulse amplitude
(cannot be used in VDD operating mode)
Setting the atrial sensitivity
(cannot be used in D00 operating mode)
Starts the atrial burst function
Securing Reocor D to patient, bed or infusion stand
Setting the AV delay
Selection of the atrial burst function
Selection of the pacing operating mode and off switch
Setting the pacing rate
Bold labels of the operating devices indicate safe values for
the intended use of the device.