Technical Safety
The external pacemaker Reocor D meets the international
standards for the safety of electro-medical devices according
to IEC 60601-1 and IEC 60601-1-2, as well as the interna-
tional standard IEC 60601-2-31 for temporary, external
pacemakers.
The following special features offer safety for the patient:
No metal parts that can be touched, according to the
definition of the IEC.
The design meets the standard for the device class CF
(cardiac floating) and is approved for direct treatment of
the heart. The pacemaker complies with the require-
ments for defibrillation protection stipulated in the inter-
national standards.
The closed protective cover protects the pacemaker
against dripping water.
WARNING!
The temporary leads that are connected to Reocor D repre-
sent a low-impedance conductor to the myocardium for
electric current. Therefore line-powered devices that are
operated in the patient's vicinity must be grounded in accor-
dance with established guidelines.
The pacemaker must not be used in areas at risk for explo-
sion.
All additional maintenance work and repairs should only be
performed by BIOTRONIK.