Warning
Acoustic signal for 2 s
Fast sequence of sounds
Acoustic signal and flashing
of the Pace and Sense LEDs
Low battery LED flashes.
a)
When using the battery type MN 1604 Duracell
Visual and Acoustic Signals
During the self-test after switching on Reocor D, all LEDs
light up and brief acoustic signals can be heard. The self-
test is completed after a few seconds.
If the self-test does not find any errors, the LEDs and
warning signals turn off.
When the self-test finds a defect, all LEDs flash continu-
ously and warning signals sound.
A required battery replacement is indicated by the flash-
ing red Low battery LED.
The Sense (green) LEDs signal sensing of a P wave or
R wave.
The Pace (yellow) LEDs signal pulse delivery.
The LEDs and acoustic signals also provide the following
warnings during operation:
Meaning
A pulse amplitude of < 1 V or a rate
of > 180 ppm is programmed.
Impedance outside of the permis-
sible range
High rate protection has been
triggered; self-test failed.
ERI has been reached.
Error correction
Check whether the set values
are suitable for the patient.
Check whether all connectors
are securely plugged in.
Check whether the leads have
the desired position.
Turn the device off and return
it to BIOTRONIK.
Replace the battery; about 36
a)
hours
®
®
, Procell
of service time remain.