Starting up
Perform calibration
Fig. 84: Software setting: 'Calibration'
[Calibration] - start:
1.
[Main menu] Press the
ð The [Main menu] is displayed.
2.
[Calibration] - select.
ð The screen for starting [Calibration] is displayed.
3.
[Start] - press.
ð Screen changes to the 'calibration step 1' .
Calibration step 1 runs automatically.
The screen page for 'calibration step 1' is called up.
4.
Enter the quantity in litres (before the decimal point) using the scrollbar.
5.
Select the quantity in millilitres after the decimal point.
6.
Enter the quantity in millilitres (after the decimal point) using the scrollbar.
7.
[Press the Next button]
ð Screen changes to the 'calibration step 2' .
The 'calibration step 2' runs automatically.
The screen page for calibration step 2 is called up.
8.
Enter the quantity in litres (before the decimal point) using the scrollbar.
9.
Select the quantity in millilitres after the decimal point.
10. Enter the quantity in millilitres (after the decimal point) using the scrollbar.
11. [Press the Next button]
ð The setting is saved, and the screen switches back to [Operating display] .