Error and Warning Messages
Error and Warning Messages
EN
Gas temperature >42°C/ Disconnect
luer lock / Cool down tube
The device terminates gas heating
and insufflation.
Error message/Call service
Device temperature error/Turn
device off
The insufflation is stopped or cannot
be started.
Venting valve defective
RTP defective
RTP deactivated
Tube set not connected
Check/change tube set
88
Cause
The temperature of the gas
exceeds 42° C.
The device does not work prop-
erly and activates the internal
safety system.
The device temperature is
above 70 °C.
The device temperature is
below 10° C.
Venting system malfunctioning. • Use the ON/OFF key to turn the device off and turn
Device defective
Device check was unsuccessful
The tube set has not been con-
nected or has not been inserted
correctly but the START function
field has been pressed.
Leak in tube connection
Troubleshooting
• Unplug the insufflation tube at the trocar.
• Press the START/STOP function field. The device
insufflates without heating the gas.
• Let hot gas escape until the tube is only warm to
the touch.
• Continue the operation without gas heating or use
a different gas heating tube.
Turn device off and on again. This reactivates the
gas heating.
• Check gas heating after surgery using a different
tube.
If this error message is displayed again, have the
device checked by a qualified service technician.
It is possible to continue using the device without gas
heating.
• Switch the device off and back on after approx. 3
seconds have expired using the ON/OFF key.
• Should the error message appear again, the device
is defective. Make sure the device can no longer be
used until a qualified service technician conducts
the appropriate tests and repairs.
• Prepare for changing the gas bottle.
• Turn the device off for about 10 minutes using the
ON/OFF key. The surrounding room temperature
must be above 10° C.
device back on after approx. 3 seconds. If this error
message is displayed again after the device check
has concluded, have the device checked by an
authorized service technician. The device can be
still operated but without venting system. The error
message is repeated with each Start/Stop.
• It is possible to continue with the surgery but do
not connect the measuring line of the RTP system.
The message is repeated with each Start/Stop.
After surgery is finished, call a service technician to
repair the device.
• Surgery can be continued but without the continu-
ous pressure measurement. Please make sure that
neither of the two tubes is connected to a trocar or
closed off during a device check. Restart device if
needed.
• Please insert a tube set or check the tube set con-
nection.
• Please insert a new tube set or check the tube set
connection.