Idiomas disponibles

Idiomas disponibles

vice check.
4. Select Vessel Harvest operating mode in the insufflation mode selection.
5. Connect an original insufflation tube set High Flow 2 (6) to the device.
6. Press the function field for the gas consumption display (4) to set the display
to 0 Liter.
7. Select the nominal pressure 10 mm Hg (2) and the gas flow 1 l/min (5).
8. Close the end of the tube (Luer Lock) (7).
9. Start insufflation: Press the START function field (3).
10. Insufflate for approx. 30 s. The display status line depicts Occlusion after max.
4 seconds.
11. Stop insufflation: Press the STOP function field (3).
WARNING!
If the actual gas consumption is higher than 0.4 l, there is a leak in the system.
12. Replace the tube set and repeat the test from step 6 to step 11.
13. If another leak is detected, this leak is then directly within the device. Make
sure the device can no longer be operated until a qualified service technician
conducts the appropriate tests and repairs.
WARNING!
Never work with a leaky insufflation tube, accessory, and/or device. This can
lead to an incorrect measurement of the actual pressure values, which can cause
an uncontrolled pressure increase in the abdomen.
The functionality of the gas heating is indicated by displaying the gas heating
symbol on the display when using the "Heated tube set with RTP for Pneumo
Sure".
9.5.1

Filling Tube System with CO2

At least 1 l of CO2 has to be let out from the system before every surgery while
the tube is connected and the end of the tube is open. This will expel any air with-
in the tube system and the device.
1. Start insufflation: Press the START function field.
Wait until the gas consumption display shows 1.0 Liter.
2. Stop insufflation: Press the STOP function field.
3. Press the gas consumption display field to reset the gas consumption display
to 0.0 Liter. This ensures the gas consumption is correctly displayed during
surgery.
The function test is complete. The device is tested and ready to be used for sur-
gery.
WARNING!
Do not use this device if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects on the power plug
and power cable.
9.6

Using the Device during Surgery

WARNING!
The function test must be performed prior to each surgery.
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Vessel Harvest Operating Mode
Testing gas heating functionality
Using the device during surgery
EN
55