Using and Controlling the PNEUMO SURE High Flow Insufflator in Pediatric Operating Mode
EN
"Real-Time Pressure Sensing" functionality
(RTP)

Using the device during surgery

40
symbol on the display when using the "Heated tube set with RTP for Pneumo
Sure".
1. Connect the corresponding tube set ("Heated with Real-Time Pressure Sens-
ing" or "High Flow with Real-Time Pressure Sensing").
2. Insufflation is initially always started intermittently. Availability of the "Real-
Time Pressure Sensing" functionality is checked automatically. If this is the
case, the device switches to a continuous mode. This is indicated by the cor-
responding symbol on the display.
7.5.1

Filling Tube System with CO2

At least 1 l of CO2 has to be let out from the system before every surgery while
the tube is connected and the end of the tube is open. This will expel any air with-
in the tube system and the device.
1. Start insufflation: Press the START function field.
Wait until the gas consumption display shows 1.0 Liter.
2. Stop insufflation: Press the STOP function field.
3. Press the gas consumption display field to reset the gas consumption display
to 0.0 Liter. This ensures the gas consumption is correctly displayed during
surgery.
The function test is complete. The device is tested and ready to be used for sur-
gery.
WARNING!
Do not use this device if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects on the power plug
and power cable.
7.6
Using the Device during Surgery
WARNING!
The function test must be performed prior to each surgery.
CAUTION!
The venting system is automatically triggered (not during Veress mode and not
if the venting valve has been deactivated in the Configuration menu) if the mea-
sured actual pressure exceeds the set nominal pressure during insufflation. The
device interrupts insufflation and releases gas until the actual pressure has
dropped below the set nominal value.
1. The device is turned on.
2. An insufflation tube set is connected.
7.6.1

Insufflating with Veress Cannula

On delivery from the factory, following values are set for Pediatric operating
mode:
• a gas flow value of 0.1 l/min and
• Veress insufflation for insufflation through the Veress cannula.
CAUTION!
Please note that with flow values up to 1 l/min the Veress insufflation operating
mode is automatically activated (the venting valve is switched off). The gas