Connecting A Gas Bottle; Connecting To Central Gas Supply - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

CO
2
4.1.1

Connecting a Gas Bottle

CAUTION!
Always use a high-pressure tube to connect gas bottle and device.
The gas bottle must be in a vertical position. The gas bottle pressure may not ex-
ceed 80 bar or be less than 15 bar.
CAUTION!
Gas bottles with riser pipe can release dirt and oily fluids into the device. Do not
use a gas bottle with riser pipe.
4.1.2

Connecting to Central Gas Supply

Use the following device connectors available as additional equipment to con-
nect to a central gas supply (house supply):
• 0620-040-003 for NIST house gas supply or
• 0620-040-002 for DISS house gas supply.
1. Attach the high-pressure tube to the gas connection.
2. Fix the high-pressure tube with the nut.
3. Tighten the nut.
The type of corresponding gas supply must be set in the configuration menu (see
chapter 10.1.3 Setting the Gas Supply Type, page 62).
Initial Device Startup
EN
15
Tabla de contenido
loading

Tabla de contenido