Configuration Menu (Overview)
EN
Configuration menu II
64
10.2
Configuration menu II
While insufflation is stopped, press the Actual function field for 2 seconds to ac-
cess the Configuration menu. In the Configuration menu I, press the to access
the Configuration menu II.
10.2.1 Setting the Gas Flow Rates*
Rate
Operating mode
Rate 1
Bariatric
High Flow
Vessel Harvest
Pediatric
Rate 2
Bariatric
High Flow
Vessel Harvest
Pediatric
Rate 3
Bariatric
High Flow
Vessel Harvest
Pediatric
*Inactive with Pediatric mode
In the configuration menu II, press the Gas flow rates function field to access the
selection menu.
The display switches to the gas supply selection menu.
For example, press the function field (Rate 1: 3 l/min) to set gas flow rate 1.
The display switches to the setting menu for gas flow rate 1.
Press the function field º or » to specify the value.
You now have the opportunity to
1. save the selected value by pressing the (SAVE) function field. The display de-
picts (SAVED) for 2 seconds. After saving you return automatically to the pre-
(2)
vious level,
2. press the function field to return to the previous menu level without sav-
(1)
ing,
3. or press (EXIT) to return to the work screen without saving.
(3)
10.2.2 Setting First Nominal Gas Flow*
*Selectable only in Pediatric mode
In the Configuration menu I, press the to access the Configuration menu II.
Factory setting
Range
5 l/min
1 -20 l/min
3 l/min
1 -20 l/min
1 l/min
1 -10 l/min
Not available
25 l/min
Rate 1 -45 l/min
20 l/min
Rate 1 -40 l/min
4 l/min
Rate 1 -10 l/min
Not available
45 l/min
Rate 2 -45 l/min
40 l/min
Rate 2 -40 l/min
10 l/min
Rate 2 -10 l/min
Not available