Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Bariatric Operating Mode
EN
50
rate. Compare the abdominal filling rate to the selected nominal pressure.
You can change the nominal gas flow and the nominal pressure during sur-
gery without interrupting the insufflation process. The Luer Lock connection
of the measuring tube can be connected to another trocar that also inserted
when using a tube set with measuring line. This enables the continuous mea-
suring of the pressure.
8.6.3
"Real-Time Pressure Sensing" functionality (RTP)
To use the "Real-Time Pressure Sensing" (RTP) please use a tube set with measur-
ing line. The Luer Lock connection of the measuring tube has to be connected to
another trocar placed during the surgery. This enables the continuous measuring
of the pressure (see 5.5 Connecting Insufflation Tube Set, page 19 and 5.6.1 Using
the direct pressure measurement function (Real-Time Pressure Sensing RTP),
page 21).
CAUTION!
Make sure the stopcock (valve) of the trocar is fully opened and that both tube
connections are connected to different trocars.
8.7

Stop the Insufflation

1. Press the STOP function field. The following values are displayed:
• Gas consumption display: last measured value
• Actual pressure: current measured value
• Actual gas flow: 0 l/min
• Nominal pressure: last set value In case of exceeding the safety limit, the
nominal pressure value will be reset to the safety limit value.
• Nominal gas flow: last set value
The status field of the display depicts Insufflation stopped followed by
Mode: Bariatric alternating with Push
2. Remove the tube set from the device. Observe applicable hygiene regulations
when disposing of the tube set.
CAUTION!
If the tube set remains connected to the device, there is the danger that leftover
fluid in the tube or the instruments will be sucked into the device.
3. Close the gas supply.
4. Use the ON/OFF switch to turn the device off.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.
to release.