Configuration Menu - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Configuration menu II
Menu level
1. Submenu level
Gas flow rate*
High Flow:

Rate 1= 3 l/min (range 1-20)
Rate 2= 20 l/min (range rate 1-40)
Rate 3= 40 l/min (range rate 2-40)
Venting con-
Bariatric:
trols
Rate 1= 5 l/min (range 1-20)

Rate 2= 25 l/min (range rate 1-45)
Rate 3= 45 l/min (range rate 2-45)
Vessel Harvest:
Rate 1= 1 l/min (range 1-10)
Rate 2= 4 l/min (range rate 1-10)
Rate 3= 10 l/min (range rate 2-10)
First nominal
Pediatric:
gas flow
0.1 l/min (setting range 0.1-1 l/min)
Max. nominal
High Flow: 30 mm Hg (setting range 5-30 mm Hg)
pressure
Bariatric: 30 mm Hg (setting range 5-30 mm Hg)

Vessel Harvest: 20 mm Hg (setting range 5-
20 mm Hg)
Pediatric: 20 mm Hg (setting range 5-20 mm Hg)
Flow safety
Limit ON
limit
Limit OFF
Warning signal
Vessel Harvest:
Occlusion
Bariatric:

Pediatric:
High Flow:
*Inactive menu items are displayed in gray.
Menu item Gas Flow Rates is inactive in Pediatric mode.
Menu item First Nominal Gas Flow is active exclusively in Pediatric mode.
Menu item Flow Safety Limit is active only in Pediatric operating mode and Vessel Harvest operating mode.
Signal ON, Signal OFF
Signal ON, Signal OFF
Signal ON, Signal OFF
Signal ON, Signal OFF
Configuration Menu (Overview)
2. Submenu level
3. Submenu
level
EN
59
Tabla de contenido
loading

Tabla de contenido