Using and Controlling the PNEUMO SURE High Flow Insufflator in Pediatric Operating Mode
EN
(1)
(2)
36
7.2
Selecting Pediatric Operating Mode
1. Selection of the insufflation mode can be done only while insufflation is
stopped. If Pediatric operating mode is not yet set and displayed, press the
function field Change Mode for 2 seconds to get the insufflation mode over-
view.
2. The display depicts an insufflation modes overview. Select the desired insuf-
flation operating mode by pressing the Pediatric function field.
3. The procedure profile is depicted on the display. The displayed parameters
correspond with the factory settings or the values set in the Configuration
menu (see chapter 10 Configuration Menu (Overview), page 58).
7.3
Presetting Nominal Pressure in Pediatric Operating Mode
The nominal pressure can be set during insufflation or while insufflation is
stopped. Values may range from 1 to max. 20 mm Hg or the value set in the Con-
figuration menu.
Increasing/decreasing nominal pressure:
Briefly press the º or » function field to increase or decrease the pressure. Keep-
ing the º or » function field depressed longer than 1.5 seconds activates scrolling
in increments of 1.
Select a nominal pressure value between 1 and 20 mm Hg. The preset value is in-
dicated in the nominal pressure display.
Safety limit:
When increasing the nominal pressure, the status line of the display depicts Safe-
ty limit starting at 12 mm Hg and the nominal value flashes.
The nominal pressure value 12 mm Hg is a limit value and should not be exceed-
ed for newborns if at all possible. Pressing the nominal pressure º function field
again does not increase the pressure any further.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
Release the function field at that point. The display switches back to the nominal
value setting after 2 seconds. Now you can set a value up to 15 mm Hg. The nom-
inal pressure 15 mm Hg is a limit value. This is where the recommended range for