Stop The Insufflation - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6.6.4

Stop the Insufflation

1. Press the STOP function field. The following values are displayed:
• Gas consumption display: last measured value
• Actual pressure: current measured value
• Actual gas flow: 0 l/min
• Nominal pressure: last set value In case of exceeding the safety limit, the
nominal pressure value will be reset to the safety limit value.
• Nominal gas flow: last set value
The status field of the display depicts Insufflation stopped followed by Mode:
High Flow alternating with Push
2. Remove the tube set from the device. Observe applicable hygiene regulations
when disposing of the tube set.
CAUTION!
If the tube set remains connected to the device, there is the danger that leftover
fluid in the tube or the instruments will be sucked into the device.
3. Close the gas supply.
4. Use the ON/OFF switch to turn the device off.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.
Using and Controlling the PNEUMO SURE High Flow Insufflator in High Flow Mode
to release.
EN