In the configuration menu I, press the First nominal pressure function field to ac-
cess the setting.
The display switches to the First nominal pressure menu.
Press the function field º or » to set the First nominal pressure.
You now have the opportunity to
1. save the selected value by pressing the (SAVE) function field. The display de-
picts (SAVED) for 2 seconds. After saving you return automatically to the pre-
vious level,
2. press the function field to return to the previous menu level without sav-
ing,
3. or press (EXIT) to return to the work screen without saving.
10.1.2 Setting the Venting Controls
In the configuration menu I, press the Venting controls function field to access
the venting system selection.
The display switches to the Venting controls selection menu.
Factory Settings
Operating
Factory setting
mode
Bariatric
With veress insufflation OFF With veress insufflation OFF,
High Flow
With veress insufflation OFF With veress insufflation OFF,
Vessel Harvest
Venting system OFF
Pediatric
With veress insufflation ON
For example, tap the Venting valve status function field.
The display switches to the Venting valve status selection menu.
With Veress Insufflation ON is only selectable while in Pediatric operating mode
and is otherwise depicted in gray.
Venting system ON is only selectable while in Vessel Harvest mode and is other-
wise depicted in gray.
Range
Venting system OFF
Venting system OFF
Venting system ON,
Venting system OFF
With Veress insufflation ON,
With Veress insufflation OFF,
Venting system OFF
Configuration Menu (Overview)
Setting the venting valve status
EN
(2)
(1)
(3)
61