Operating the Device - General
Removing the tube set
EN
Connecting the gas heating
Fig. 5-4
Connecting the gas heating
Plug for insufflation tube with gas
(1)
heating
(2)
Internal heating
Insufflation tube
(3)
Measuring tube
(4)
Gas heating display
Set Pressure
Actual Pressure
mm Hg
15
Gas heater OK
Mode
STOP
High Flow
20
Press the plug of the insufflation tube in the direction of the device. This releases
the snap-in latch and you are now able to remove the insufflation tube set.
• 2 long acoustic warning signal are emitted,
• the message Tube set not connected is displayed,
5.6
Use the heating tube to insufflate lukewarm gas (37° C).
1. Switch device on.
2. Connect an insufflation tube with gas heating (Heated tube set with Real-
The heating module is located directly in the insufflation tube.
CAUTION!
Do not subject the heater tube to direct heat (e.g., operating room lamp, en-
doscop with light source) or high room temperatures.
The device automatically determines whether a tube set with or without gas
heating is connected. The display depicts a symbol
Set Flow
Gas heater OK after the successful detection of the corresponding tube set.
mm Hg
l/min
0
3
0.0
Liter
Actual
3
l/min
Using the Gas Heating
Time Pressure Sensing (RTP).
(4)
(1)
(2)
(3)
and the status line reads