flation tube.
7. Stop insufflation: Press the STOP function field.
The following values are displayed:
Actual pressure 0.0 [mm Hg]
Gas consumption >0.0 [l]
8. Press the gas volume function field.
Gas consumption >0.0 [l]
The basic function test of the device is complete.
13.3
Testing the Pressure Sensors in High Flow Operating Mode
(1)
mm Hg
(4)
1. Set to High Flow operating mode.
2. Select a nominal gas flow rate of 1.0 l/min.
Do not press the START/STOP function field.
CAUTION!
Never use the syringe to extract gas from the device.
3. Connect a manometer (1) and an air-filled syringe (4) to the insufflation
tube connection (3)/adapter (2).
4. Use the syringe to generate a pressure of at least 10 mm Hg, which registers
on the manometer.
Actual pressure display: 10 ±2 [mm Hg]
5. Use the syringe to generate a pressure of at least 20 mm Hg, which registers
on the manometer.
Actual pressure display: 20 ±2 [mm Hg]
6. Use the syringe to generate a pressure of at least 30 mm Hg, which registers
on the manometer.
Actual pressure display: 30 ±2 [mm Hg]
(3)
(2)
Annual Inspection
EN
79