Performing The Function Test In Bariatric Operating Mode Before Using The Device During Surgery - Stryker PNEUMO SURE Instrucciones De Uso

Ocultar thumbs Ver también para PNEUMO SURE:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Insufflation: Bariatric
While Insufflation: Bariatric any pressure loss due to leaks can be quickly equal-
ized. The APC Technology (Advanced Pressure Control) enables raising the actual
pressure gently to the level of the nominal pressure. In case of large volumes, the
actual pressure does not exceed the nominal pressure (see chapter 11 Safety
functions, page 70).
Start the device by pressing the START function field.
With a nominal flow setting of < 6 l/min Veress insufflation is displayed. Above
5 l/min Insufflation: Bariatric is displayed.
Stop the device by pressing the STOP function field.
NOTE!
Tube, filter and instrument can reduce the gas flow.
8.5
Performing the Function Test in Bariatric Operating Mode before
Using the Device during Surgery
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
For your own safety and that of your patient, use only original accessories.
WARNING!
The function test must be performed prior to each surgical procedure.
(1) (2) (3)
(8)
(7)
1. The device is switched off, no tube set is connected.
2. Make sure the gas supply is connected and open.
3. Use the ON/OFF switch (1) to turn the device on. The device conducts a de-
Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Bariatric Operating Mode
(4)
(5)
(6)
Preparation
Testing the device
Fig. 8-1
Layout for testing the device
ON/OFF switch
(1)
Nominal pressure display
(2)
START/STOP function field
(3)
(4)
Gas consumption display
Nominal flow display
(5)
Insufflation tube set
(6)
Stopcock (valve)
(7)
Veress cannula
(8)
EN
47