Using and Controlling the PNEUMO SURE XL High Flow Insufflator in Vessel Harvest Operating Mode
EN
Preparation
Testing the device
Fig. 9-1
Layout for testing the device
ON/OFF switch
(1)
Nominal pressure display
(2)
START/STOP function field
(3)
(4)
Gas consumption display
Nominal flow display
(5)
Insufflation tube set
(6)
Luer Lock connection
(7)
54
Briefly press the nominal gas flow function field » to reduce the nominal gas
flow.
Safety limit:
The safety limit can be activated or deactivated in the Configuration menu.
When increasing the nominal flow, the status line of the display depicts Safety
limit starting at 6 l/min and the nominal value flashes.
Release the function field to increase the nominal gas flow further. The display
switches back to the nominal value setting after 2 seconds. Now you can set a
value of up to 10 l/min.
CAUTION!
Exceeding this safety limit is to be decided by and the responsibility of the user/
operator.
9.5
Performing the Function Test in Vessel Harvest Operating Mode
before Using the Device during Surgery
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
For your own safety and that of your patient, use only original accessories.
WARNING!
The function test must be performed prior to each surgical procedure.
(1) (2) (3)
(7)
1. The device is switched off, no tube set is connected.
2. Make sure the gas supply is connected and open.
3. Use the ON/OFF switch (1) to turn the device on. The device conducts a de-
(4)
(5)
(6)