Safety and Electromagnetic Compatibility (EMC)
WARNING! This equipment is designed and tested as Class A equipment defined in GB 4824. It
may cause radio interference, in which case, you may need to take measures to mitigate the interference.
WARNING! This equipment is designed and tested as Class A equipment defined in CISPR 11.
This equipment is not intended for use in residential environments and may not provide adequate
protection to radio reception in such environments.
WARNING! It is the user's responsibility to ensure that a compatible electromagnetic
environment can be maintained in order that the equipment will perform as intended. It is recommended
that the electromagnetic environment should be evaluated prior to operation of the equipment.
WARNING! Do not use this device in proximity to sources of strong electromagnetic radiation
(e.g. unshielded intentional RF (radio frequency) sources), as these can interfere with proper operation.
WARNING! Only qualified, Sensititre trained service engineers or technicians should undertake
repair and service of the instrument.
WARNING! Emissions which exceed the levels required EMC standard can occur when the EUT
is connected to a test object. NOTE. A test object(s) used with the instrument do not contain electrical or
electromagnetic component(s).
WARNING! This equipment is designed for use in a PROFESSIONAL HEALTHCARE FACILITY
ENVIRONMENT. It is likely to perform incorrectly if used in a HOME HEALTHCARE ENVIRONMENT. If it
is suspected that performance is affected by electromagnetic interference, correct operation may be
restored by increasing the distance between the equipment and the source of the interference.
Some parts of the AIM system operate at a hazardous voltage. Only qualified, Sensititre trained
service engineers or technicians should undertake repair and service of the instrument.
Sensititre AIM™ Instructions for Use